These products have been discontinued. For comparable products, click here for the following:
- Criticare VitalCare Vital Signs Monitor w/ DOX SpO2, NIBP, HR and Printer
- Criticare VitalCare Vital Signs Monitor w/ Nellcor SpO2, NIBP, HR, Printer
- Criticare VitalCare Vital Signs Monitor w/ DOX Sp02, NIBP, TurboTemp, Printer, HR
- Criticare VitalCare Vital Signs Monitor w/ Nellcor Sp02, NIBP, TurboTemp, Printer, HR
Designed to travel everywhere you do, the VitalCareTM 506N3 series
patient monitor with integrated carry-handle attaches quickly and easily
to a rolling stand. Available in multiple configurations – with choice of
Criticare or Nellcor oximetry – the 506N3 patient monitors provide
maximum versatility and cost-effectiveness.
New smart battery management allows up to 100 NIBP measurements
on a single charging. With a compact design that is easily setup for spotcheck
or continuous monitoring, the 506N3 patient monitor adapts
readily to hospital, clinic and transport requirements.
Features
and Benefits
- Proprietary ComfortCuff ™ NIBP
measures on inflation for faster,
more comfortable readings.
- Choice of Criticare DOX™ digital
SpO2 technology or Nellcor
OxiMax™ SpO2 technology.
- Audible and visual alarms are
adjustable providing instant
notification the way you need
them.
- Optional FILAC FasTemp™
predictive temperature for quick,
accurate measurements.
- Optional internal printer for
on-demand, automatic interval,
trends or multi-patient reports.
Technical Specifications
Non-Invasive Blood Pressure
- Technique:
Oscillometric measure upon inflation
- Measurement Time: <40
seconds typical; 0 to 300mmHg;
- standard adult cuff
- Automatic
Measurement
- Cycles: 1 - 60 minute; 2 - 4 hour; STAT
- Inflation
Pressure Range: 30 to 300mmHg (Adult)
- 30 to 150mmHg (Pediatric)
- 20
to 150mmHg (Neonate)
- Resolution: 1mmHg
- Clinical Accuracy:
SP-10:2002
- Transducer Accuracy: ±2mmHg or ±2% of reading, whichever
is
- greater, over full range
- STAT mode: 5 minutes of consecutive
readings
Pulse Oximetry
- Range:
1% to 99%
- Resolution: 1%
- Accuracy (DOX only): 70% to 99% range:
±2%;
- 50% to 69% range: ±3%;
- <50%: unspecified
- Accuracy
(OxiMax Only): 70% to 99% range: ±2%;
- <70%: unspecified
- Method:
Dual wavelength LED
- Audio: Tone with each detected pulse; pitch
varies with saturation
Heart
(Pulse) Rate
- Range: 20 to 300bpm
- Accuracy (DOX only):
Greater of ±1bpm or ±1%, of reading
- Unspecified for <30 or >240
bpm
- Update Time: Every second
- Resolution:
1bpm
- Pulse Volume: 10 levels or Off
Temperature (optional)
- Type: FILAC FasTemp™
- Method:
Predictive or Continuous
- Channels: 1
- Range: 30.0º to 43.0º C
- Accuracy
Continuous Mode:±0.2°F (±0.1°C)
- Accuracy Predictive Mode: ±0.2°F
(±0.1°C) for ranges 37°C to 39 °C
- (Compliant with EN 12470-3 and
- ASTM
E1112:98)
- Resolution: 0.1º F or 0.1º C
- Predictive Response Time:
Oral: 10 seconds typical;
- Axial: 10 seconds typical;
- Rectal: 10
seconds typical
- Probes/Wells: Oral/Axial FILAC FasTemp [202026]
- Rectal
FILAC FasTemp (optional) [202034]
- Isolation Well, Oral, Axial
[202089]
- Isolation Well, Rectal (optional) [202087]
Alarms
- Indication: Audible; Visual
- Volume:
10 levels or Off
- Silence: Yes; 2 minutes or permanent
Trend Memory
- Memory: Last 24 hours
- Parameters:
SpO2, pulse rate, NIBP, temperature
- Report Types: Single Patient or
Multi-patient;
- Selectable formats
Displays
- Type: LED
- Parameters: NIBP, SpO2;
temperature, pulse rate; pulse bar
- Status: Alarm Silence; Sensor; Low
Battery; External
- Power; Patient Size
- Text: Alphanumeric LCD
- Languages:
English, French, German, Italian, Portuguese, Spanish, Dutch, Danish
System Output
- Com Port: Serial, RS
232-compatible
Printer
- Type:
Thermal array
- Tabular Mode: 1 - 60 minute intervals; 2 - 24 hour
intervals;
- Auto on BP; Auto on Alarm, Off
Mechanical
- Weight: Base unit with
battery: 6.95 lb; 3.15kg
- Size: Base unit: 8.7" (H) x 7.1" (W) x 6.5”
(D); 22cm (H) x 18cm (W) x 16.5cm (D)
Power Requirements
- Voltage: 100 to
240 VAC, 50 to 60 Hz
- Battery: Sealed lead acid; Life: 8 hours
- Rechargeable:
6 hours
Environmental
- Operating
Temperature: 32º to 113º F; 0º to 45º C
- Storage Temperature: 23º to
122º F; -5º to 50º C
- Operating Humidity: 15% to 90%, non-condensing
Classification
- Type of Protection:
Class I equipment
- Degree of Protection: Type CF defibrillator-proof
equipment
- Protection against Ingress: IPX1 rating, Drip-Proof
Equipment
Warranty Information
|